Abstract
While first generation porous coatings have had clinical success, aseptic loosening remains a leading cause of revision. The purpose of this study was to investigate the reasons for revision and to assess the amount of bone ingrowth in retrieved porous tantalum components. In a prospective multicenter retrieval program, 76 porous tantalum acetabular shells, 5 femoral stems, 7 patellas and 36 tibial trays were collected from revision surgeries. A subset of the implants was analyzed for bone ingrowth. The main reason for revision was infection for acetabular shells (1.4. years implantation time) and instability for tibial trays (1.8. years implantation time). Two of the thirty primary surgery acetabular shells and one of the thirty-six primary surgery tibial trays were revised for implant loosening. We observed full depth penetration of bone into the porous tantalum layer for the acetabular shells and femoral stems.
| Original language | English |
|---|---|
| Pages (from-to) | 922-927 |
| Number of pages | 6 |
| Journal | Journal of Arthroplasty |
| Volume | 28 |
| Issue number | 6 |
| DOIs | |
| State | Published - Jun 2013 |
Keywords
- Aseptic loosening
- Bone ingrowth
- Porous tantalum components
Fingerprint
Dive into the research topics of 'Bone ingrowth in well-fixed retrieved porous tantalum implants'. Together they form a unique fingerprint.Cite this
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver