Electromagnetic interference from a muscle stimulation device causing discharge of an implantable cardioverter defibrillator: Epicardial bipolar and endocardial bipolar sensing circuits are compared

  • Taya V. Glotzer
  • , Margaret Gordon
  • , Mark Sparta
  • , Glauco Radoslovich
  • , John Zimmerman
  • HACKENSACK UNIVERSITY MEDICAL CENTER

Research output: Contribution to journal › Article › peer-review

34 Scopus citations

Abstract

This case report is about two patients with two different types of ICDs who underwent electrical muscle stimulation (EMS) therapy. In one patient with an ICD that has epicardial screw-in bipolar sensing leads, electromagnetic interference (EMI) from the EMS device caused the delivery of an inappropriate ICD discharge. In a second patient with an ICD with endocardial true bipolar sensing, there was no evidence of EMI during the EMS therapy despite all of our attempts to reproduce it. The sensing circuits in the two different ICDs are compared.

Original languageEnglish
Pages (from-to)1996-1998
Number of pages3
JournalPACE - Pacing and Clinical Electrophysiology
Volume21
Issue number10
DOIs
StatePublished - 1998

Keywords

  • Electrical muscle stimulation
  • Electromagnetic interference
  • Implantable cardioverter defibrillator