Skip to main navigation Skip to search Skip to main content

Fretting and Corrosion Damage in Taper Adapter Sleeves for Ceramic Heads: A Retrieval Study

  • Daniel W. MacDonald
  • , Antonia F. Chen
  • , Gwo Chin Lee
  • , Gregg R. Klein
  • , Michael A. Mont
  • , Steven M. Kurtz
  • , Harold E. Cates
  • , Matthew J. Kraay
  • , Clare M. Rimnac
  • Drexel University
  • Rothman Orthopaedic Institute
  • University of Pennsylvania
  • Cleveland Clinic Foundation
  • Exponent, Inc.
  • Tennessee Orthopaedic Clinics
  • Case Western Reserve University

Research output: Contribution to journal › Article › peer-review

23 Scopus citations

Abstract

Background During revision surgery with a well-fixed stem, a titanium sleeve can be used in conjunction with a ceramic head to achieve better stress distribution across the taper surface. In vitro testing suggests that corrosion is not a concern in sleeved ceramic heads; however, little is known about the in vivo fretting corrosion of the sleeves. The purpose of this study was to investigate fretting corrosion in sleeved ceramic heads in retrieved total hip arthroplasties. Methods Thirty-seven sleeved ceramic heads were collected during revision. The femoral heads and sleeves were implanted 0.0-3.3 years. The implants were revised predominantly for instability, infection, and loosening. Fifty percent of the retrievals were implanted during a primary surgery. Fretting corrosion was assessed using the Goldberg-Higgs semiquantitative scoring system. Results Mild-to-moderate fretting corrosion scores (score = 2-3) were observed in 92% of internal tapers, 19% of external tapers, and 78% of the stems. Severe fretting corrosion was observed in 1 stem trunnion that was previously retained during revision surgery and none of the retrieved sleeves. There was no difference in corrosion damage of sleeves used in primary or revision surgery. Conclusion The fretting corrosion scores in this study were predominantly mild and lower than reported fretting scores of cobalt-chrome heads in metal-on-polyethylene bearings. Although intended for use in revisions, we found that the short-term in vivo corrosion behavior of the sleeves was similar in both primary and revision surgery applications. From an in vivo corrosion perspective, sleeves are a reasonable solution for restoring the stem taper during revision surgery.

Original languageEnglish
Pages (from-to)2887-2891
Number of pages5
JournalJournal of Arthroplasty
Volume32
Issue number9
DOIs
StatePublished - Sep 2017

Keywords

  • femoral stem
  • fretting corrosion
  • revision hip arthroplasty
  • sleeved ceramic head
  • taper corrosion